The table below shows the intra-subject (scan-rescan) CoV of MT, PD, R1 and R2* values in the white matter at different isotropic resolutions (2.7 mm, 2.0 mm, 1.6 mm, 1.3 mm, 1.0 mm). All quantitative maps were corrected for B1+ inhomogeneities using an acquired B1+ map. The intra-subject CoV with and without correcting for Gibb’s ringing is presented for each quantitative map and resolution.

Voxel size (mm isotropic) PD mean (%) MT mean (%) R1 mean (%) R2* mean (%)
original unringing original unringing original unringing original unringing
2.7 0.85 0.78 1.94 1.97 1.53 1.65 3.74 3.52
2.0 0.91 0.83 2.09 2.01 1.59 1.58 3.61 3.39
1.6 1.11 0.95 2.96 2.71 1.64 1.53 5.01 4.38
1.3 1.46 1.27 2.88 2.58 2.14 1.96 5.41 4.77
1.0 2.04 1.78 3.88 3.47 2.96 2.68 7.95 7.01